High resolution small animal single photon emission computed tomography: uptake of [99mTc]bombesin and
F Scopinaro1, R Massari, A D Varvarigou
1Department of Nuclear Medicine, La Sapienza University of Rome, Rome, Italy. francesco.scopinaro@uniroma1.it
Aim:
The aims of this study were: 1) to perform brain single photon emission computed tomography (SPECT) in anesthetized rats with high resolution cameras (HRC) equipped with parallel hole collimation resolution of about 1 mm (HRC1) and 2 mm (HRC2); 2) to assess when and with which radio-tracer HRC1 SPECT shows advantages over HRC2.
Methods:
We used two multicrystal HRCs with parallel square hole collimators, whose pure tungsten septa closely fit the crystals, in turn matched with a 4 inch2 position sensitive photomultiplier. HRC1 showed 1.1 mm and HCR2 2.1 mm resolution at collimator contact. HRCs performed 180 degrees semi-circular orbits around the head of rats: image reconstruction occurred with ordered subsets expectation maximization algorithms. Resolution of SPECT was measured with a Derenzo Phantom, resulting 1.4 mm for HRC1 and 2.3 mm for HRC2. Three rats were studied with [(99m)Tc]HMPAO, 3 rats with [(99m)Tc]bombesin (BN) and 48 h later with [(123)I]ioflupane (DaTSCAN). SPECT studies were reviewed by two experienced operators.
Results:
Technetium-99m-HMPAO SPECT showed similar images with HRC1 and HRC2. The uptake of BN by amygdale, hippocampus and olfactory tract was detected by both cameras. DaTSCAN SPECT with HRC1 showed detailed image of the tail of the caudatus: this image was not obtained with HRC2. DaTSCAN and BN SPECT showed amygdale with both HRCs. However, only the central nucleus of amygdale takes up DaTSCAN, whereas central, lateral and basolateral amygdaloid nuclei express BN receptors. Only HRC1 SPECT showed amygdale larger with BN than with DaTSCAN.
Conclusion:
Spatial resolution of 1.4 mm is appropriate to detect selected subcerebral structures.
More Related Videos
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
08:40Positron Emission Tomography Imaging for In Vivo Measuring of Myelin Content in the Lysolecithin Rat Model of Multiple Sclerosis
Published on: February 28, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...