Separation of polar mushroom toxins by mixed-mode hydrophilic and ionic interaction liquid
Wai-cheung Chung1, Sau-ching Tso, Sai-tim Sze
1Government Laboratory, 7/F Homantin Government Offices, 88 Chung Hau Street, Homantin, Hong Kong Special Administrative Region. swcchung@govtlab.gov.hk
Abstract:
Reversed-phase liquid chromatography (RPLC) is commonly used to analyze nonvolatile contaminants and naturally occurring toxins in foods. However, polar compounds, such as hydrophilic polypeptides and quaternary ammonium salts, are often not satisfactorily separated by RPLC and present a challenge for analytical scientists. In this study, hydrophilic interaction liquid chromatography (HILIC), on an amide-based stationary phase in combination with electrospray ionization (ESI) tandem mass spectrometry (MS-MS), is successfully employed to simultaneously separate polar mushroom toxins, including amanitins and phallotoxins, which are cyclic oligopeptides and muscarine, a quaternary ammonium compound, in mushrooms. The sensitivity of different ionization modes is studied, and the positive ionization mode is found to provide a more sensitive and effective tool for the unambiguous identification of the concerned polar toxins because of their characteristic fragmentation patterns. The properties of the mobile phase are also found to have significant impacts on the separation. At a high acetonitrile (ACN) concentration, hydrophilic interaction dominates, and all analytes under study demonstrate a much higher affinity with the stationary phase. The addition of methanol (MeOH) as a modifier could further enhance the HILIC separation for amanitins, phallotoxins, and muscarine. Valley-to-valley separation is achieved upon the optimatizatiqn of the mobile phase (comprising of ACN, MeOH, and ammonium formate buffer at pH approximately 3.5) and the solvent gradient. HILIC coupled with ESI-MS-MS is demonstrated to be a novel technique for the simultaneous separation and confirmatory analysis of the concerned polar toxins by providing an environment of solubility and retention that could not be achieved through the use of RPLC.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
09:36Applications of Liquid-Chromatography Tandem Mass Spectrometry in Natural Products Research: Tropane Alkaloids as a Case Study
Published on: March 8, 2024
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...