Evaluation of repeatability for the automatic estimation of endothelial cell density in donor corneas
Alfredo Ruggeri1, Enrico Grisan, Jan Schroeter
1Department of Information Engineering, University of Padova, Padova, Italy. alfredo.ruggeri@unipd.it
Background/Aims:
The repeatability of an automatic system for estimation of endothelial cell density (ECD) from microscopy images in donor corneas was assessed.
Methods:
A computer program for the automatic ECD estimation from frequency analysis was previously presented. An evaluation of its repeatability was performed on a data set containing 200 corneas by assessing the accuracy and precision of automatic versus manual values. For each cornea, 2-21 images (1536 total) at 100x for automatic ECD and one image at 200x for manual ECD were available.
Results:
Accuracy of automatic ECDs was -45 (SD 99) cells/mm(2) (-2% (4%)). Precision of repeated automatic ECDs on the same cornea was 62 (30) cells/mm(2) (3% (1%)). The algorithm was also capable of identifying all non-processable images.
Conclusion:
The proposed automatic technique proved to be reliable for its routine use in a typical eye bank setting like the one considered here.