An investigation of the bacterial contamination of small animal breathing systems during routine use
Ludovic Pelligand1, Richard Hammond, Andrew Rycroft
1Anaesthesia Service, The Royal Veterinary College, Hertfordshire, UK. lpelligand@rvc.ac.uk
Objective:
To investigate the need for sterilization of anaesthetic breathing systems to prevent cross-infection between animals due to the re-use of anaesthetic circuit tubing.
Study Design:
Prospective microbiological study.
Methods:
Bacteriology samples were taken from 37 sterile breathing systems, each used for 1 day, at two sampling sites (one proximal and one distal to the animal). The number of patient connections, cumulative anaesthesia time, culture results, number of colony-forming units and the number of different species were recorded. Secondly, four sterile breathing systems were used for 2 months under routine conditions and sampled every 2 weeks and the same parameters recorded. Finally, the inner surfaces of four sterile breathing systems were inoculated with a known load of canine oropharyngeal bacteria. Bacteriology samples were taken at 1 minute, 1 hour and 1 day post-deposition. The number of colonies identified was compared with the initial load.
Results:
Only a very small number of micro-organisms were isolated and these were generally organisms of low pathogenic potential. The proximal site was found to be significantly more colonized than the distal site (p < 0.001). Neither longer daily connection time (p = 0.54), nor a higher number of connections (p = 0.81) increased the incidence of proximal site colonization. Over the 2-month study period, the bacterial population did not increase. There was no correlation between cultures isolated from successive samples taken from the same tubing. There was rapid loss of viability of the micro-organisms deliberately inoculated onto the tubing surface: the number of colonies isolated from the breathing system after 1 minute was significantly lower than in the inoculum (p = 0.042).
Conclusions And Clinical Relevance:
Sterile anaesthesia breathing systems were colonized by environmental micro-organisms of low pathogenicity. Although long-term survival of recognized pathogens in a dry environment is still possible, the use of a bacterial filter or a sterilized breathing system for routine veterinary anaesthesia cannot be supported by current evidence.
More Related Videos
04:32Murine Oropharyngeal Aspiration Model of Ventilator-associated and Hospital-acquired Bacterial Pneumonia
Published on: June 28, 2018
06:11Rapid Detection of Bacterial Pathogens Causing Lower Respiratory Tract Infections via Microfluidic-Chip-Based Loop-Mediated Isothermal Amplification
Published on: March 29, 2024
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Microbiota of the Respiratory Tract
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...