Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Noninvasive cardiac output monitoring (NICOM): a clinical validation
Pierre Squara1, Dominique Denjean2, Philippe Estagnasie2
1CERIC, Clinique Ambroise Paré, 27, boulevard Victor Hugo, 92200, Neuilly-sur-Seine, France. pierre.squara@wanadoo.fr.
Objective:
To evaluate the clinical utility of a new device for continuous noninvasive cardiac output monitoring (NICOM) based on chest bio-reactance compared with cardiac output measured semi-continuously by thermodilution using a pulmonary artery catheter (PAC-CCO).
Design:
Prospective, single-center study.
Setting:
Intensive care unit.
Patients:
Consecutive adult patients immediately after cardiac surgery.
Interventions:
Cardiac output measurements obtained from NICOM and thermodilution were simultaneously recorded minute by minute and compared in 110 patients. We evaluated the accuracy, precision, responsiveness, and reliability of NICOM for detecting cardiac output changes. Tolerance for each of these parameters was specified prospectively.
Measurements And Results:
A total of 65,888 pairs of cardiac output measurements were collected. Mean reference values for cardiac output ranged from 2.79 to 9.27 l/min. During periods of stable PAC-CCO (slope<+/-10%, 2SD/mean<20%), the correlation between NICOM and thermodilution was R=0.82; bias was +0.16+/-0.52 l/min (+4.0+/-11.3%), and relative error was 9.1%+/-7.8%. In 85% of patients the relative error was <20%. During periods of increasing output, slopes were similar with the two methods in 96% of patients and intra-class correlation was positive in 96%. Corresponding values during periods of decreasing output were 90% and 84%, respectively. Precision was always better with NICOM than with thermodilution. During hemodynamic challenges, changes were 3.1+/-3.8 min faster with NICOM (p<0.01) and amplitude of changes did not differ significantly. Finally, sensitivity of the NICOM for detecting significant directional changes was 93% and specificity was 93%.
Conclusion:
Cardiac output measured by NICOM had most often acceptable accuracy, precision, and responsiveness in a wide range of circulatory situations.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...