Purpose:
To examine the association between scanning laser polarimetry (SLP), using enhanced (ECC) and variable corneal compensation (VCC) with optical coherence tomography (OCT), and to compare their discriminating ability in the diagnosis of glaucoma.
Methods:
Normal and glaucomatous eyes enrolled from four clinical sites underwent complete examination, automated perimetry, SLP-ECC, SLP-VCC, and OCT. Eyes were characterized in two groups based on the typical scan score (TSS): Normal birefringence pattern (NBP) was defined as a TSS of 80 to 100 and abnormal birefringence pattern (ABP) as TSS
Results:
Ninety-five normal volunteers and 63 patients with glaucoma were enrolled. Average visual field mean deviation was -4.2 +/- 4.3 dB in the glaucoma group. In eyes with NBP, SLP-ECC had significantly (all P
Conclusions:
Compared with SLP-VCC, SLP-ECC has significantly stronger correlations with OCT and may improve the discriminating ability for early glaucoma diagnosis.
More Related Videos
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy (oSLO) and Optical Coherence Tomography (OCT)
Published on: August 4, 2018
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment