Effects of CT irradiation on implantable cardiac rhythm management devices
Cynthia H McCollough1, Jie Zhang, Andrew N Primak
1Department of Radiology, Mayo Clinic College of Medicine, 200 First St SW, Rochester, MN 55905, USA. mccollough.cynthia@mayo.edu
Purpose:
To prospectively measure the response of a variety of models of implantable cardiac rhythm management devices (ICRMDs) to the radiation delivered by computed tomography (CT), for both maximum and typical dose levels.
Materials And Methods:
Twenty-one ICRMDs (13 pacemakers, eight cardioverter-defibrillators) manufactured by Medtronic (Minneapolis, Minn) were exposed to ionizing radiation from CT systems in both spiral and dynamic acquisition modes at maximum and typical dose levels. Devices were monitored during exposure to check for any operational abnormalities and were interrogated after exposure to check for any residual abnormalities. Total radiation dose and peak dose rate were measured, and the volume CT dose index was recorded.
Results:
Oversensing was observed in 20 of 21 devices at maximum doses and in 17 of 20 devices at typical doses. Oversensing most often manifested as inhibition, although it occasionally manifested as tracking or safety pacing. Two devices inhibited for more than 4 seconds in spiral mode at clinical dose levels. Oversensing was transient and ceased as soon as the device stopped moving through the x-ray beam or the beam was turned off. The partial electrical reset (PER) safety feature was activated in two models, InSync 8040 and Thera DR. With the exception of PER, programming was not altered. Effects occurred only if the x-ray beam passed directly over the ICRMD.
Conclusion:
CT irradiation at typical clinical doses results in oversensing of ICRMDs in the majority of devices tested, although the identified effects were predominantly transient.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Dysrhythmias VI: Management of Dysrhythmias
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...