Purpose:
To assess current tonometer disinfection practice in the UK, and compare with published recommendations.
Methods:
Every ophthalmology unit with training recognition in the UK was contacted (n=155). A senior nurse at each institution completed a telephone questionnaire regarding local tonometer disinfection practice.
Results:
The response rate was 100%. Thirty-five units (23%) reported exclusive use of disposable tonometer heads and were excluded from further analysis. One hundred and twenty units (77%) used either reusable or a combination of reusable and disposable tonometer heads. Where reusable heads were used, 80 units (67%) immersed them in a chlorine-based solution such as sodium hypochlorite or sodium dichloroisocyanurate. Others used isopropyl alcohol (18 units), hydrogen peroxide (12 units), chloramine (5 units), chlorhexidine (4 units) and peracetic acid (1 unit). Where a chlorine-based agent was used, the concentration of available chlorine ranged from 125 to 30 000 p.p.m., with 50 units (63%) using a concentration of less than 5 000 p.p.m. (i.e., inadequate based on published recommendations). Where the tonometer head was immersed in disinfectant between patients (n=101), 29 units (29%) provided just one tonometer head per practitioner, making adequate soak time between patients unlikely. Every unit replenished the disinfectant at least daily, deemed sufficient for most agents. However, hydrogen peroxide solutions should be replenished twice daily, which did not take place in nine units.
Conclusion:
This survey reveals disparity between current tonometer disinfection practice and published international recommendations, with some institutions using practices that may render patients susceptible to transmissible infection.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...