Method for evaluating drip-rate accuracy of intravenous flow-regulating devices.

B C Carleton1, R J Cipolle, S D Larson

  • 1Department of Pharmacy Practice, College of Pharmacy, University of Minnesota (UM), Minneapolis 55455.

American Journal of Hospital Pharmacy
|November 1, 1991
PubMed
Summary

Patient movement significantly impacts intravenous infusion accuracy. Devices like the roller clamp, Dial-A-Flo, and Exacdrop showed reduced accuracy with position changes, unlike the IVAC 280 and 3M IV Flow Regulator.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

TCERG1L allelic variation is associated with cisplatin-induced hearing loss in childhood cancer, a PanCareLIFE study.

NPJ precision oncology·2021
Same author

Review: The Perioperative Use of Thromboelastography for Liver Transplant Patients.

Transplantation proceedings·2018
Same author

Adjuvant pretreatment with alum protects neonatal mice in sepsis through myeloid cell activation.

Clinical and experimental immunology·2017
Same author

Genetic markers of cisplatin-induced hearing loss in children.

Clinical pharmacology and therapeutics·2014
Same author

Pharmacogenomic diversity in Singaporean populations and Europeans.

The pharmacogenomics journal·2014
Same author

Response to "evaluation of pharmacogenetic markers to predict the risk of Cisplatin-induced ototoxicity".

Clinical pharmacology and therapeutics·2014

Area of Science:

  • Medical Devices
  • Fluid Dynamics
  • Patient Care

Background:

  • Intravenous (IV) therapy is critical for medication delivery.
  • Accuracy of IV flow-regulating devices is essential for patient safety.
  • Patient positioning and movement can potentially affect IV drip rates.

Purpose of the Study:

  • To investigate the impact of patient movement and position changes on the drip-rate accuracy of various IV flow-regulating devices.
  • To compare the accuracy of different IV devices under dynamic conditions.

Main Methods:

  • Twenty healthy adult volunteers participated in the study.
  • IV infusion sites were established, and 5% dextrose was administered using standard IV tubing and bags.
  • Drip rates were measured at 40 drops/min for five devices (IVAC 280, roller clamp, Dial-A-Flo, Exacdrop, 3M IV Flow Regulator) across supine, sitting, and standing positions, and during walking.

Related Experiment Videos

Main Results:

  • The IVAC 280 and 3M IV Flow Regulator maintained consistent drip rates across position changes.
  • The roller clamp, Dial-A-Flo, and Exacdrop showed significant decreases in drip rate accuracy when changing from supine to sitting and sitting to standing.
  • While walking had minimal impact, returning to a supine position significantly increased drip rates for the roller clamp, Dial-A-Flo, and Exacdrop.

Conclusions:

  • The drip-rate accuracy of the roller clamp, Dial-A-Flo, and Exacdrop devices is significantly affected by patient position changes.
  • The IVAC 280 and 3M IV Flow Regulator demonstrate superior stability in drip rate accuracy during patient movement.
  • Clinical settings should consider the limitations of certain IV flow regulators when patient mobility is a factor.