Cleveland Clinic PediPump lamb cadaver fitting studies
Related Experiment Video
Updated: Jul 15, 2026

C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
Faruk Cingoz1, Kiyotaka Fukamachi, Yoshio Ootaki
1Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, OH, USA.
Abstract:
We are developing the PediPump, a magnetically suspended, mixed-flow pump, as an implantable pediatric ventricular assist device (VAD). Lamb cadaver fitting studies were performed to determine the optimal pump location and optimal design of the inflow and outflow conduits for chronic in vivo studies. A prototype of the PediPump right and left ventricular assist devices (RVAD and LVAD, respectively) were implanted via a sternotomy or left thoracotomy in four lamb cadavers (3.7-34.2 kg). Via a sternotomy, the RVAD and LVAD required long inflow cannulas when placed into the right or left thoracic cavities, respectively. Via a left thoracotomy, with both pumps implanted in the left thoracic cavity, the RVAD required a 70 degrees inflow cannula and a bent outflow graft while the LVAD required a 130 degrees inflow cannula and a straight outflow graft. In conclusion, left thoracotomy provided optimal fitting for both LVAD and RVAD for the small lamb model.