High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Cine CT for attenuation correction in cardiac PET/CT
Adam M Alessio1, Steve Kohlmyer, Kelley Branch
1Department of Radiology, University of Washington, Seattle, WA 98195-7987, USA. aalessio@u.washington.edu
Respiratory motion during PET/CT scans causes artifacts. Cine CT, acquiring multiple low-dose CT images, improves attenuation correction for cardiac PET by reducing motion-related errors, with the intensity-maximum image being more robust.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Radiology
Background:
- Positron emission tomography/computed tomography (PET/CT) systems use CT for PET attenuation correction.
- Respiratory motion during PET scans can cause misalignment between CT-derived attenuation maps and PET emission data.
- This misalignment leads to errors in attenuation correction factors and artifacts in the final PET images.
Purpose of the Study:
- To evaluate the efficacy of cine CT in improving attenuation correction for cardiac PET studies.
- To compare cine CT methods (average and intensity-maximum images) against conventional helical CT for mitigating respiratory motion artifacts.
Main Methods:
- Acquired cine CT and cardiac PET data from a cardiac phantom and patient studies.
- Evaluated conventional helical CT, average cine CT, and intensity-maximum cine CT for attenuation correction.
- Assessed image alignment and quality through visual evaluation by three readers in 74 patient studies.
Main Results:
- Helical CT-based attenuation correction introduced significant artifacts in phantom studies.
- Cine CT methods reduced these artifacts, with intensity-maximum cine CT showing a 60% reduction.
- Patient studies demonstrated acceptable alignment for 77% of average cine CT and 88% of intensity-maximum cine CT cases.
Conclusions:
- Cine CT is a viable alternative to helical CT for compensating respiratory motion in cardiac PET attenuation correction.
- Both average and intensity-maximum cine CT images reduce respiration-induced misalignment errors.
- Intensity-maximum cine CT appears to be a more reliable method than the average cine CT for improved attenuation correction.