Mechanical characteristics of locking and compression plate constructs applied dorsally to distal radius fractures
Scott Boswell1, Terence E McIff, Cory A Trease
1Department of Orthopedic Surgery, University of Kansas Medical Center, Kansas City, KS 66160, USA.
Purpose:
Locking plates are thought to have many advantages such as a decreased incidence of loss of reduction secondary to screw toggling and improved bone healing due to an increased periosteal blood supply. We hypothesized that locking plates will also provide increased stiffness and increased load to failure when they are applied dorsally to stabilize dorsally comminuted distal radius fractures. This study compared the stiffness and strength of dorsally applied locking and standard (nonlocking) T-plates applied to a dorsally comminuted distal radius fracture model.
Methods:
Sixteen pairs of embalmed cadaveric human radii were potted, and a standard wedge osteotomy was performed simulating a dorsally comminuted distal radius fracture. The radii were randomized into 2 groups, so that 8 pairs received a 3.5-mm dorsal locking T-plate over the osteotomy on the right radius and 8 pairs received the same on the left radius. A dorsal 3.5-mm standard T-plate was placed over the osteotomy on the contralateral radius in each group. An axial load was used to test the strength and stiffness of each construct. Paired t tests were then used to compare the strength and stiffness of the locking plate with those of the standard plate.
Results:
A significant difference was found in both the stiffness and the strength between the locking and standard nonlocking plates. The locking T-plate was 33% stiffer than the standard T-plate. The locking T-plate had a 91% increase in the load to failure. Failure for both locking and standard T-plates occurred via volar cortex bone fracture.
Conclusions:
Locking T-plates increased both the stiffness and strength of dorsally comminuted distal radius fractures compared with standard nonlocking T-plates by a statistically significant margin.
More Related Videos
08:39Fabrication of Compressed Hosiery and Measurement of its Pressure Characteristic Exerted on the Lower Limbs
Published on: May 27, 2020
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Degree of Curvature and Radius of Curvature
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Radius of Gyration of an Area
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...
Schwarzschild Radius and Event Horizon
The minimum speed required to launch a projectile from the surface of an object to which it is gravitationally bound so that it eventually escapes the object’s gravitational field is called the escape velocity. The escape velocity is independent of the mass of the object. Merging the idea of escape...