[Definition of maximum permissible errors for major physical parameters affecting the result of microplate ABO-RH1
O Bouix1, P Sanchez, J-M Reifenberg
1EFS Pyrénées-Méditerrannée, GHU Caremeau, place Pr.-R.-Debré, CS 68223, 30942 Nimes cedex 9, France. Olivier.bouix@efs.sante.fr
Aim:
Metrological survey of analytic equipment is developing well in French blood bank laboratories, but still the limits of critical parameters affecting the result of basic methods are usually not defined. The aim of this study is to establish experimentally maximum permissible errors for critical physical values involved in microplate ABO-RH1 blood grouping, such as temperature and time of incubation and volumes of reagents and blood samples.
Materials And Methods:
Samples were internal quality controls used routinely in the laboratory. Three set points were defined for each parameter: incubation temperature 18, 22 and 26 degrees C; incubation time 10, 15 and 20 min; reagent sera and sample cells volume 20, 30 and 40 microL; reagent cells volume 15, 20 et 25 microL; plasma sample volume 30, 40 et 50 microL. All equipments (temperature test chamber, chronometer and pipettors) were calibrated using national standards (COFRAC) according to appropriate methods for estimation of uncertainty of measurement. Significant components of uncertainty were taken into account to define a global uncertainty for each measurable quantity.
Results:
All tests results from any combination of the above set points (n=81 for each sample tested in triplicate) were consistent with the expected result. Not one undetermined result was observed.
Conclusion:
The experimental design of this study allowed us to define accurately maximum permissible errors for major critical parameters involved in microplate ABO-RH1 blood grouping. Hence these values must be applied to metrological survey of the equipment. Further, our study demonstrates the robustness of the method.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Margin of Error
The ABO Blood Group
Antigens in the ABO Blood Group System
Antigens are substances that can trigger an immune response, leading to the production of antibodies. In the ABO blood group system,...
Random and Systematic Errors
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...