Study of mechanical degradation of UHMWPE acetabular components due to clinical X-ray procedures
Related Experiment Video
Updated: Jul 15, 2026

Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Letícia A Vasconcellos1, Eduardo Blando, André A Souto
1Gepsi--Centro de Pesquisa e Desenvolvimento Em Física--Tecnopuc, Pontifícia Universidade Católica do Rio Grande do Sul, P.O. Box 1429, 90619-900 Porto Alegre, RS, Brazil.
Abstract:
The use of multi-component femoral implants to replace the femur head and re-establish bone motion has been widespread since the 70s. Frequently these implants have spherical metallic heads made of, for example, 316-L stainless steel or Cr-Co alloys, which allow rotational motion towards a polymeric component (UHMWPE). One of the major causes of implant rejection is the generation of UHMWPE debris on the surface between the implant head and the polymeric component. The gamma ray sterilization of implants and the periodical X-ray medical control could contribute to premature degradation of the polymeric surface, resulting in increased wear and shortened lifetime of the implant. In this work we study the degradation degree of the polymeric UHMWPE component as function of the X-ray dose. The elasto-plastic deformation and recovery were carried out by means of a nanohardness tester equipment and the polymer degradation was measured using a fast Fourier transform infra-red (FT-IR) equipment. The results show the compromise among the irradiation doses, the surface oxidation and the mechanical properties of the samples.