Related Experiment Video
Updated: Jul 15, 2026

Thermal Measurement Techniques in Analytical Microfluidic Devices
Published on: June 3, 2015
Application of calorimetry on a chip to high-pressure materials
Alexandra Navrotsky1, Maria Dorogova, Frances Hellman
1Thermochemistry Facility and Nanomaterials in the Environment, Agriculture, and Technology Organized Research Unit, University of California, Davis, CA 95616, USA. anavrotsky@ucdavis.edu
Abstract:
Silicon micromachined calorimeters ("calorimeter on a chip") are used to measure heat capacities and phase transition enthalpies for thin film, single crystal, and powder samples (5-500 mug). The technology is thus compatible with the small samples produced in multianvil and large diamond anvil cells. Techniques for handling small samples and attaching them to the calorimetric devices have been developed. Initial data illustrate application to CoO and to Fe(2)SiO(4) olivine and spinel, a quenched high pressure phase metastable at ambient conditions. The calorimetric entropy of the olivine-spinel transition in Fe(2)SiO(4) (-16 +/- 5 J/mol.K) is in good agreement with that calculated from phase equilibrium data (-14 +/- 3 J/mol.K). A magnetic transition in iron silicate spinel, detected previously by Mossbauer spectroscopy, is seen in the calorimetric signal.