Impedance cardiography: a useful and reliable tool in optimization of cardiac resynchronization devices
Konstantin M Heinroth1, Marcel Elster, Sebastian Nuding
1Department of Medicine III/Cardiology, Martin-Luther-University Halle-Wittenberg, E, rnst-Grube-Strasse 40, 06097, Halle (Saale), Germany. konstantin.heinroth@medizin.uni-halle.de
Aims:
Optimizing cardiac resynchronization therapy (CRT) devices has become more complex since modification of both atrioventricular (AV) and interventricular (VV) stimulation intervals has become possible. The current paper presents data from the routine use of impedance cardiography (IC)-based cardiac output (CO) measurements to guide the optimization of AV- and VV-interval timing of CRT devices.
Methods And Results:
Forty-six patients with heart failure (left ventricular ejection fraction <35%, New York Heart Association (NYHA) III-IV) and left bundle branch block (>130 ms) in sinus rhythm were evaluated 3-5 days after implantation of a CRT device by means of IC. CO was measured without pacing and with biventricular pacing using a standard protocol of VV- and AV-interval modification from -60 to +60 ms and 80 to 140 ms, respectively, in 20 ms steps. Mean CO without pacing was 3.66 +/- 0.85 L/min and significantly increased to 4.40 +/- 1.1 L/min (P<0.05) with simultaneous biventricular pacing and an AV interval of 120 ms. 'Optimizing' both VV and AV intervals further increased CO to 4.86 +/- 1.1 L/min (P<0.05). Maximum CO was measured in most patients with left ventricular pre-excitation. The proportion of 'non-responders' to CRT was reduced by 56% following AV- and VV-interval modification using IC guidance.
Conclusion:
Modification of both AV and VV intervals in patients with a CRT device significantly improves CO compared with standard simultaneous biventricular pacing and no pacing. IC is a useful non-invasive technique for guiding this modification. Marked variability of optimal AV and VV intervals between patients requires optimization of these intervals for each patient individually.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:25Impedance Pneumography for Minimally Invasive Measurement of Heart Rate in Late Stage Invertebrates
Published on: April 4, 2020
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Bode Plots Construction
Cardiomyopathy IV: Restrictive Cardiomyopathy
Cardiomyopathy II: Dilated Cardiomyopathy