The trouble with CTD100.

John M Boone1

  • 1Department of Radiology, University of California Davis Medical Center, Ellison Building, 4860 Y Street, Suite 3100, Sacramento, California 95817, USA. jmboone@ucdavis.edu

Medical Physics
|May 16, 2007
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The end of browsing, continuous publication, and title scrutiny.

Medical physics·2026
Same author

Assessing variations in 3D image quality in chest CT across sites and scanners.

Medical physics·2026
Same author

Multi-x-ray source array for stationary tomosynthesis or multi-cone angle cone beam CT.

Proceedings of SPIE--the International Society for Optical Engineering·2026
Same author

Hybrid simulation of breast CT for assessing microcalcification detectability.

Journal of medical imaging (Bellingham, Wash.)·2025
Same author

Medical Physics is transitioning to double blinded review.

Medical physics·2025
Same author

Empirical Motion-Artifact Reduction for Non-Rigid Motion in Dedicated Breast CT.

IEEE transactions on bio-medical engineering·2025
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

The computed tomography dose index (CTDI100) remains valid for wider CT scanner beams, but its overall efficiency is low, potentially impacting dose accuracy even with narrower beams.

Area of Science:

  • Medical Physics
  • Radiological Dosimetry

Background:

  • Computed tomography dose index (CTDI100) is a standard metric for CT radiation dose measurement.
  • Its efficiency with modern multi-slice CT scanners and wide x-ray beams is not well understood.

Purpose of the Study:

  • To assess the efficiency of CTDI100 for wider collimated x-ray beams using Monte Carlo simulations.
  • To evaluate the impact of beam width on CTDI100 accuracy.

Main Methods:

  • Monte Carlo simulations of a CT scanner geometry with polyenergetic x-ray spectra.
  • Computation of dose spread functions (DSFs) in polymethyl methacrylate (PMMA) phantoms.
  • Calculation of CTDI100 efficiency for various slice thicknesses (1-400 mm).

Main Results:

Related Experiment Videos

  • CTDI100 efficiency was found to be 82-90% for 10 mm beams in head phantoms and 63-88% in body phantoms.
  • Efficiency decreased minimally (by ~1%) for 40 mm beams compared to 10 mm beams.
  • The study highlights that CTDI100 efficiency is inherently low even for narrower beams.

Conclusions:

  • CTDI100 remains a valid metric for wider x-ray beams (up to 40 mm) in CT scanners.
  • The inherent low efficiency of CTDI100 measurement may limit the accuracy of derived dose computations.
  • Further research into dose metrics for improved accuracy in CT dosimetry is warranted.