Related Experiment Video
Updated: Jul 14, 2026

Measurement of Bioelectric Current with a Vibrating Probe
Published on: January 4, 2011
Lock-in amplifier error prediction and correction in frequency sweep measurements
Maximiliano Osvaldo Sonnaillon1, Fabian Jose Bonetto
1Laboratorio de Cavitación y Biotecnología, San Carlos de Bariloche, Argentina. msonnaillon@ing.uncr.edu.ar
Abstract:
This article proposes an analytical algorithm for predicting errors in lock-in amplifiers (LIAs) working with time-varying reference frequency. Furthermore, a simple method for correcting such errors is presented. The reference frequency can be swept in order to measure the frequency response of a system within a given spectrum. The continuous variation of the reference frequency produces a measurement error that depends on three factors: the sweep speed, the LIA low-pass filters, and the frequency response of the measured system. The proposed error prediction algorithm is based on the final value theorem of the Laplace transform. The correction method uses a double-sweep measurement. A mathematical analysis is presented and validated with computational simulations and experimental measurements.
Related Concept Videos
Small-Signal Analysis of MOSFET Amplifiers
Frequency Response of Op Amp Circuits
Frequency Response and Gain:
The gain of the op-amp, A(ω), is not a constant but a function of the input signal frequency. An op-amp can maintain a constant gain at low frequencies, known...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
MOSFET Amplifiers
Inverting and Non-inverting OpAmps
Small-Signal Analysis of BJT Amplifiers