Addressing Practical Issues in Atomic Force Microscopy-Based Micro-Indentation on Human Articular Cartilage Explants
Published on: October 28, 2022
Cross-talk correction in atomic force microscopy
A Hoffmann1, T Jungk, E Soergel
1Institute of Physics, University of Bonn, Bonn, Germany.
Abstract:
Commercial atomic force microscopes usually use a position-sensitive photodiode to detect the motion of the cantilever via laser beam deflection. This readout technique makes it possible to measure bending and torsion of the cantilever separately. A slight angle between the orientation of the photodiode and the plane of the readout laser beam, however, causes false signals in both readout channels. This cross-talk may lead to misinterpretation of the acquired data. We demonstrate this fault with images recorded in contact mode on periodically poled ferroelectric crystals and present a simple electronic circuit to compensate for it. This circuit can correct for cross-talk with a bandwidth of approximately 1 MHz suppressing the the false signal to <<1%.