Imaging of radiation-sensitive samples in transmission electron microscopes equipped with Zernike phase plates
Marek Malac1, Marco Beleggia, Ray Egerton
1National Institute for Nanotechnology, 11421 Saskatchewan Drive, Edmonton, Canada T6G 2M9. mmalac@phys.ualberta.ca
Abstract:
We have optimized a bright-field transmission electron microscope for imaging of high-resolution radiation-sensitive materials by calculating the imaging dose n(0) needed to obtain a signal-to-noise ratio (SNR)=5. Installing a Zernike phase plate (ZP) decreases the dose needed to detect single atoms by as much as a factor of two at 300 kV. For imaging larger objects, such as Gaussian objects with full-width at half-maximum larger than 0.15 nm, ZP appears more efficient in reducing the imaging dose than correcting for spherical aberration. The imaging dose n(0) does not decrease with extending of chromatic resolution limit by reducing chromatic aberration, using high accelerating potential (U(0)=300 kV), because the image contrast increases slower than the reciprocal of detection radius. However, reducing chromatic aberration would allow accelerating potential to be reduced leading to imaging doses below 10 e(-)/A(2) for a single iodine atom when a CS-corrector and a ZP are used together. Our simulations indicate that, in addition to microscope hardware, optimization is heavily dependent on the nature of the specimen under investigation.
More Related Videos
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
09:17Scanning Transmission Electron Microscopy Tomography in Virology: 3D Imaging of High-pressure Frozen, Freeze-substituted Samples
Published on: August 6, 2025
Related Concept Videos
Transmission Electron Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Preparation of Samples for Electron Microscopy
Overview of Electron Microscopy
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...