Evaluation of an automated soluble transferrin receptor (sTfR) assay on the Roche Hitachi analyzer and its comparison
Christine M Pfeiffer1, James D Cook, Zuguo Mei
1Centers for Disease Control and Prevention, National Center for Environmental Health, Atlanta, GA 30345, USA. cpfeiffer@cdc.gov
Background:
Soluble transferrin receptor (sTfR) assays are currently not standardized. This hinders data comparison between studies and also affects the use of a recently proposed model to estimate body iron.
Methods:
We evaluated the analytical performance of a fully automated sTfR immunoturbidimetric assay (Roche Diagnostics) and compared it with two ELISA assays (Ramco Laboratories and an in-house ELISA assay used in the body iron model).
Results:
The Roche assay showed excellent intra- and inter-assay precision (CV<5%). Prolonged exposure of serum samples to room temperature and multiple freeze-thaw cycles did not affect sTfR concentrations. Receiver-operator characteristic curve analysis demonstrated that the Roche assay (area-under-the-curve (AUC)=0.882) was superior to the Ramco assay (AUC=0.794) in predicting iron deficiency (defined as serum ferritin <10 microg/L; P=0.013). Method comparison between the Roche and the two ELISA assays showed good correlations (r>0.8); however, sTfR values by the Roche assay were on average 30% lower than values obtained with the two ELISA assays.
Conclusions:
sTfR data measured with an immunoturbidimetric assay can be compared to a commonly used ELISA assay, and can be used in the body iron model through regression equations obtained in the present study.