Measurements of the neutron dose near a 15 Mv medical linear accelerator
N Golnik1, M Zielczynski, W Bulski
1Institute of Precision and Biomedical Engineering, Warsaw University of Technology, Warsaw, Poland. golnik@mchtr.pw.edu.pl
Abstract:
In this work, simplified recombination methods for routine estimation of dose equivalent in mixed (gamma and neutrons) radiation field outside the irradiation field of linear medical accelerators is considered. The author's earlier reported method of H(10) measurements, involving determination of the recombination index of radiation quality, Q(4) by tissue-equivalent recombination chamber was combined with the new method for determination of the photon to neutron dose ratio D(X)/D(n) from the ratio of ion collection efficiencies measured in the investigated radiation field and in two reference fields of gamma and neutron radiations. The method is suitable when the neutron contribution to the total absorbed dose, D(n)/D, is >3%.