Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Optimisation of the readout parameters when evaluating thermal neutron doses by TL dosimetry with LiF:Mg,Ti
U German1, M Weinstein, A Abraham
1Nuclear Research Center Negev, P.O. Box 9001, 84190 Beer Sheva, Israel. gur118@inter.net.il
Abstract:
The location of the glow peaks depends on the heating rate. It takes some time until the crystal reaches the heater temperature, and this time lag has a direct effect on the shift of peaks towards higher temperatures. Some information on the high-temperature peaks may be lost if the readout conditions (mainly length of time) are not properly chosen. Step heating profiles to a varying final temperature between 300 degrees C and 125 degrees C were used to study the time dependence of the collected information in the glow curves of (6)LiF:Mg,Ti crystals, and the minimal heating time for evaluation of thermal neutron doses was determined.