Does long-term continuous transcranial Doppler monitoring require a pause for safer use?
Kiyotaka Nakagawa1, Toshihiro Ishibashi, Masato Matsushima
1Department of Anesthesiology, ME Laboratory, Research Center for Medical Science, Jikei University School of Medicine, Tokyo, Japan. k-n@jikei.ac.jp
Cerebrovascular Diseases (Basel, Switzerland)
|May 24, 2007
Summary
Long-term Transcranial Doppler sonography (TCD) monitoring can increase brain temperature. To prevent thermal damage, researchers suggest pausing TCD every 30 minutes during continuous use.
Area of Science:
- Neuroscience
- Medical Imaging
- Biophysics
Background:
- Transcranial Doppler sonography (TCD) is a common tool for monitoring cerebral blood flow.
- Potential thermal effects of prolonged TCD insonation on brain tissue have been under-investigated.
- Ultrasound absorption by the skull and brain tissue can lead to temperature increases.
Purpose of the Study:
- To quantify the actual temperature rise in brain tissue during long-term TCD insonation.
- To determine if pauses are necessary for safe, continuous TCD monitoring.
- To establish safety guidelines for extended TCD applications.
Main Methods:
- In vivo animal study using 15 New Zealand White rabbits.
- Thermocouples inserted at the skull-brain interface (SBI) to measure temperature changes.
- TCD applied at maximum output (0.2 W, 2 MHz) for 90 minutes, with a control group.
Main Results:
- Skull-brain interface temperature (SBIT) in the TCD group increased by 3.47°C within 25 minutes, reaching a plateau.
- The estimated maximum safe duration for continuous TCD application was determined to be 33 minutes.
- Significant temperature elevation indicates a thermal risk associated with prolonged TCD use.
Conclusions:
- Cerebral blood flow per brain weight, a key factor in heat dissipation, shows similarities between rabbits and humans.
- Findings suggest that clinical long-term TCD monitoring should incorporate a pause after approximately 30 minutes of insonation.
- Implementing intermittent TCD application is recommended to mitigate the risk of thermal brain damage.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
Pulse rhythm
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...