Investigation of coil phase compensation in 3D imaging at very high acceleration factors
Mary Preston McDougall1, Steven M Wright
1Department of Electrical Engineering, Texas A&M University, 3120 TAMU, College Station, TX 77843, USA. mpmcdougall@tamu.edu
Purpose:
To investigate the confounding effect of the coil phase in highly accelerated parallel imaging with small coils, contextualize the effect in terms of single-echo acquisition (SEA) imaging, and show that it can be managed in the case of 3D imaging.
Materials And Methods:
The effects of the coil phase variations in a 64-channel array of surface microcoils were modeled. Fully encoded 64 x 128 x 64 (N(phase enc) x N(readout) x N(slice enc)) 3D data sets were obtained, from which factor of 64 accelerated 3D image sets (1 x 128 x 64 each) were extracted from single phase-encoding lines, each representing a different phase compensation value.
Results:
A comparison of the SEA images indicates that the choice of a compromise value for phase compensation successfully enabled a straightforward extension of SEA imaging to three dimensions. The use of the single compromise compensation value in the 3D acquisition resulted in a signal-to-noise ratio (SNR) penalty ranging from 6% to 41% through the slab when compared to the highest SNR possible using any phase compensation value.
Conclusion:
The coil-related phase shift issues inherent to highly accelerated imaging will require further study, but this work indicates the general nature of the problem and, more auspiciously, shows that it can be mitigated for at least this application.
More Related Videos
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...