Detection of dideuterium in diprotium gas at infinitesimal concentration using a trace reduction detector
Takao Kawano1, Naohiro Tsuboi, Hirotsugu Tsujii
1National Institute for Fusion Science: 322-6 Oroshi, Toki, Gifu 509-5292, Japan. kawano.takao@nifs.ac.jp
Abstract:
An analyzer able to detect extremely small concentrations of dideuterium in diprotium gas was developed. To examine its performance, four gas samples of pure diprotium with dideuterium concentrations of 0.25, 0.5, 1, and 5 cm(3)/m(3) were measured using the analyzer, and peak areas and retention times from 18 chromatograms were determined. Results indicated that the peaks were distributed within a relative standard deviation of approximately 15% and a linear correlation existed between peak area and concentration. Retention times were stable, showing a relative standard deviation of 1% with repeated tests and having no correlation with concentration of the four gas samples.
More Related Videos
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Gas Chromatography: Types of Detectors-II
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...