Validation and reproducibility assessment of tigecycline MIC determinations by Etest
Anne Bolmström1, Asa Karlsson, Anette Engelhardt
1AB BIODISK, Solna, Sweden.
Abstract:
A multicenter study was conducted to validate Etest tigecycline compared to the Clinical Laboratory Standards Institute reference broth microdilution and agar dilution methodologies. A large collection of gram-negative (n = 266) and gram-positive (n = 162) aerobic bacteria, a collection of anaerobes (n = 385), and selected collections of nonpneumococcal streptococci (n = 369), Streptococcus pneumoniae (n = 372), and Haemophilus influenzae (n = 372) were tested. Strains with reduced susceptibility to tigecycline were used with all test methods. The Etest showed excellent inter- and intralaboratory reproducibility for all organism groups tested regardless of the test methodology. The essential agreement values with the reference method (+/-1 dilution) were >99% for the collection of gram-negative and gram-positive aerobes; >98% for the S. pneumoniae, H. influenzae, and anaerobe collections; and 100% for the group of nonpneumococcal streptococci. These results validate the performance accuracy and utility of Etest tigecycline and verify the reproducibility of this convenient predefined gradient methodology for tigecycline susceptibility determination.
More Related Videos
08:01A Streamlined, Label-Free Real-Time 50% Tissue Culture Infectious Dose (TCID50) Assay using Impedance for Automated Viral Titer Quantification
Published on: January 9, 2026
10:50Design and Use of a Low Cost, Automated Morbidostat for Adaptive Evolution of Bacteria Under Antibiotic Drug Selection
Published on: September 27, 2016