Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Isokinetic muscle testing for weak patients suffering from neuromuscular disorders: a reliability study
Vincent Tiffreau1, Isabelle Ledoux, Bruno Eymard
1Institut de Myologie, GH Pitié-Salpêtrière, 75651 Paris cedex 13, France. v-tiffreau@chru-lille.fr
Neuromuscular Disorders : NMD
|June 1, 2007
Summary
This study presents a reliable isokinetic knee strength testing method for neuromuscular disorder (NMD) patients. The new technique accurately measures muscle function, correlating well with manual muscle testing (MMT).
Area of Science:
- Neurology
- Rehabilitation Medicine
- Biomechanics
Background:
- Accurate muscle strength assessment is crucial for evaluating neuromuscular disorders (NMD).
- Existing methods may lack the sensitivity required for patients with varying degrees of muscle weakness.
- Isokinetic dynamometry offers a controlled environment for objective strength measurement.
Purpose of the Study:
- To introduce and validate a novel isokinetic knee flexor and extensor testing procedure for NMD patients.
- To assess the reliability of this method in individuals with mild to severe muscle weakness.
- To investigate the correlation between the proposed isokinetic testing and traditional manual muscle testing (MMT).
Main Methods:
- Utilized the Biodex 3 system in continuous passive motion (CPM) mode for isokinetic testing.
- Calculated muscle torque by subtracting passive motion torque from active, concentric effort torque.
- Evaluated reliability through repeated testing in 15 NMD patients and compared results with MMT using correlation coefficients.
Main Results:
- High reliability was demonstrated, with intraclass correlation coefficients (ICCs) ranging from 0.91 to 0.99 for peak torque, work, and power.
- Moderate to strong correlations (Spearman rho 0.67–0.74, p<0.01) were found between isokinetic dynamometry values and MMT.
- Reliability for the angle at peak torque ranged from 0.50 to 0.90.
Conclusions:
- The described isokinetic testing method is reliable and sensitive for evaluating knee muscle strength in NMD patients.
- This dynamic assessment provides valuable, objective data for NMD patient evaluation.
- The method shows significant correlation with MMT, supporting its clinical utility.