Measuring Cardiac Autonomic Nervous System (ANS) Activity in Children
Published on: April 29, 2013
Normality of upper and lower peripheral pulse transit time of normotensive and hypertensive children
1Biomedical Engineering Research Centre, Nanyang Technological University, 50 Nanyang Drive, Research Techno Plaza, Xfrontiers Block, Singapore, Singapore. jong@ntu.edu.sg
Objectives:
The ankle-brachial index (ABI) is known to be indicative of sub-clinical peripheral arterial diseases that are correlated with cardiovascular disease risk factors like atherosclerosis or ischemic extremity. Due to its occluding measurement nature, this may not be appealing to less cooperative patients when multiple prolonged screening is required. A simple and non-intrusive approach termed pulse transit time ratio (PTTR) has recently shown to be potential surrogate marker for the prolonged ABI measurement. Other studies have also suggested that subjects with hypertension have stiffer arterial wall and thereby can confound transit time related parameters. Thus, it becomes important to understand the PTTR normality and difference of hypertensive children when compared to those measured from normotensive children.
Methods:
About 55 normotensive (39 male; aged 8.4 +/- 2.3 yr) and 4 hypertensive (4 male; aged 5-10 yr) Caucasian children were recruited from the same geographical location. A customized device was used to compute pulse transit time related measurements. Since the PTTR approach produced a delta value that was similar to that of ABI, possible inter-subject physiologic differences had limited confounding effects on the derived PTTR parameter.
Results:
The obtained transit time measurements from the hypertensive children had lower pulse transit time value when compared to their normotensive counterparts. However, the statistical analysis indicated that they had insignificant PTTR difference (p > 0.01) from those seen in the normotensive children.
Conclusions:
The findings herein suggest that stiffer arterial wall may have confounding effects on the derived transit time related measurements but it is limited on the PTTR parameter. Similar to the ABI approach, PTTR may be only confounded by abnormal local changes in either of the measured peripheral arterial wall. Hence, the PTTR technique shows promise to be an ABI marker from this perspective.
Related Concept Videos
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Special considerations while measuring pulse
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the diastolic...
Regulation of Pulse