Related Experiment Video
Updated: Jul 14, 2026

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
An impedimetric immunosensor for the detection of autoantibodies directed against gliadins
1University of Applied Sciences Wildau, Biosystems Technology, Bahnhofstr. 1, 15745 Wildau, Germany.
Abstract:
An immunosensor has been developed for the detection of autoantibodies directed against wheat gliadin, a protein fraction of cereal gluten which is involved in celiac disease. The immunosensor is based on the immobilization of gliadins onto gold electrodes covered with a polyelectrolyte layer of poly(4-styrenesulfonic acid sodium salt). The immobilization was monitored by quartz crystal microbalance (QCM) analysis. The antigen-antibody interaction signal was amplified by an incubation step with peroxidase-labeled immunoglobulins and subsequent peroxidase-catalyzed oxidation of 3-amino-9-ethylcarbazole (AEC). Changes in the insulating properties of the electrode layer were measured by electrochemical impedance spectroscopy (EIS) in the presence of ferri/ferro-cyanide. Impedance spectra could be fitted to a Randles equivalent circuit with high accuracy. Exposing the sensor electrodes to various antigliadin antibody concentrations resulted in proportional changes in the charge transfer resistance. A calibration graph for the detection of antigliadin antibodies was established for antibody concentrations between 10(-8) and 10(-6) M. Finally, the sensor was used for the determination of antigliadin autoantibodies of the IgG and IgA type in several human sera.