Related Experiment Video
Updated: Jul 14, 2026

Accumulation and Analysis of Cuprous Ions in a Copper Sulfate Plating Solution
Published on: March 20, 2019
[Flame atomic absorption spectrometric determination of copper and lead in beer after preconcentration using a rapid
Yao-dong Su1, Bing-bing Hao, Li-jun Qin
1Department of Chemistry, Tongji University, Shanghai 200092, China.
Abstract:
A method was proposed for the determination of trace copper and lead in beer with flame atomic absorption spectrometry after preconcentration of copper and lead by rapid coprecipitation technique with 8-oxyquinoline-Mg(II) using manganese as an internal standard at pH 9. The standard addition recovery of lead is between 97.6%-103.0%. The detection limit is 6.28 x 10(-3) microg x mL(-1) for copper and 2.26 x 10(-2) microg x mL(-1) for lead when the sample volume is 100 mL. The effect of matrix can be overcome by the method and the results are satisfying. The method proposed here is rapid and has good reproducibility.
Related Concept Videos
Precipitation and Co-precipitation
Flame Photometry: Lab
UV–Vis Spectroscopy: Beer–Lambert Law
Extraction: Advanced Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...