Water-equivalent dosimeter array for small-field external beam radiotherapy.

Louis Archambault1, A Sam Beddar, Luc Gingras

  • 1Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA.

Medical Physics
|June 9, 2007
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Biomarkers for Precision Prognosis in Prostate Cancer: Imaging, Molecular, and Integrated Approaches.

Cancers·2026
Same author

Evaluation of the hyperspectral multicomponent calibration for plastic scintillation dosimetry.

Physics in medicine and biology·2026
Same author

Pelvic nodes ultra-hypo fractionated versus conventionally fractionated IMRT with HDR brachytherapy in prostate cancer: interim analysis of a collaborative multi-institutional non-inferiority phase 3 trial (PCS-XI, NCT05820633).

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology·2026
Same author

Assessing inter-observer variability in prostate and GTV segmentation on mpMRI: A comparison between radiation oncologists and AI-based method.

Journal of applied clinical medical physics·2026
Same author

Robust external-beam calibration of plastic scintillation detectors and uncertainty analysis for<i>in-vivo</i>dosimetry in HDR brachytherapy.

Physics in medicine and biology·2026
Same author

AAPM task group 317 report: A joint AAPM and ESTRO report on brachytherapy catheter, needle, and applicator tracking technology.

Medical physics·2026
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

A new scintillating fiber dosimeter array offers precise, rapid dose evaluation for external beam radiotherapy. This tool accurately measures small photon fields and irregular shapes, crucial for advanced radiation treatments.

Area of Science:

  • Medical Physics
  • Radiotherapy Dosimetry
  • Radiation Detection

Background:

  • External beam radiotherapy requires advanced dosimetry due to complex dose patterns.
  • Existing dosimeters face challenges in accurately measuring small and irregular radiation fields.

Purpose of the Study:

  • To evaluate a novel dosimeter array prototype using water-equivalent plastic scintillating fibers.
  • To assess the array's capability for precise and rapid dose evaluation in radiotherapy.

Main Methods:

  • Developed a prototype scintillating fiber dosimeter array.
  • Investigated spatial resolution, angular dependence, and camera light collection.
  • Compared detector performance against a small-volume ion chamber.

Related Experiment Videos

Main Results:

  • The scintillating fiber array demonstrated no measurable angular dependence.
  • Accurate dose evaluation for small photon fields and dose profiles.
  • Agreement with ion chamber measurements, with slight discrepancies in specific small fields.

Conclusions:

  • The scintillating fiber dosimeter array is a valuable tool for external beam radiotherapy.
  • It accurately measures small, irregular fields and various incident angles.
  • Suitable for intensity-modulated radiotherapy, radiosurgery, and tomotherapy.