Related Experiment Video
Updated: Jul 14, 2026

Highly Resolved Intravital Striped-illumination Microscopy of Germinal Centers
Published on: April 9, 2014
The power of single and multibeam two-photon microscopy for high-resolution and high-speed deep tissue and intravital
Raluca Niesner1, Volker Andresen, Jens Neumann
1Helmholtz Centre for Infection Research, Junior Research Group Immunodynamics, D-38124 Braunschweig, Germany.
Abstract:
Two-photon microscopy is indispensable for deep tissue and intravital imaging. However, current technology based on single-beam point scanning has reached sensitivity and speed limits because higher performance requires higher laser power leading to sample degradation. We utilize a multifocal scanhead splitting a laser beam into a line of 64 foci, allowing sample illumination in real time at full laser power. This technology requires charge-coupled device field detection in contrast to conventional detection by photomultipliers. A comparison of the optical performance of both setups shows functional equivalence in every measurable parameter down to penetration depths of 200 microm, where most actual experiments are executed. The advantage of photomultiplier detection materializes at imaging depths >300 microm because of their better signal/noise ratio, whereas only charge-coupled devices allow real-time detection of rapid processes (here blood flow). We also find that the point-spread function of both devices strongly depends on tissue constitution and penetration depth. However, employment of a depth-corrected point-spread function allows three-dimensional deconvolution of deep-tissue data up to an image quality resembling surface detection.
Related Concept Videos
Super-resolution Fluorescence Microscopy
Three-Dimensional Microscopy in Microbiology
Two-Dimensional Microscopy in Microbiology

