Skin dose measurement for patients using imaging plates in interventional radiology procedures
Hiroko Ohuchi1, Toshimitsu Satoh, Yoichi Eguchi
1Graduate School of Pharmaceutical Sciences, Tohoku University, 6-3 Aoba, Aramaki, Aoba-ku, Sendai, Miyagi, and Yamagata University Hospital, 2-2-2 Iidanishi, Yamagata, Yamagata, 990-9585 Japan. hiroko@mail.pharm.tohoku.ac.jp
Abstract:
A method using europium-doped BaFBr imaging plates (IPs) has been developed to estimate and map values of entrance skin doses during interventional radiology (IR). IPs offer many advantages for measuring the entrance skin dose because they have a wide dynamic range (up to 100 Gy), provide high spatial resolution as a detector of two-dimensional images, and can be used repeatedly. The entrance skin dose was measured by fitting a 40x40 cm IP sheet around a patient's back using a corset in clinical studies involving IR procedures at two hospitals. The corset can minimize a geometric discrepancy in dose estimates between the IP and the patient body. The entrance skin dose was measured by using photoluminescent glass dosimeters simultaneously, and both values were compared. The spatial relative dose profiles from both dose estimates showed generally good agreement; however, the doses obtained with glass dosimeter chips were often lower than those obtained with IPs. This discrepancy comes from a radiation shielding effect for x rays by IPs and a strong angular dependence of the glass dosimeter in low energy x-ray fields. Comprehensive results of this study demonstrated that IPs were able to measure entrance skin dose in even high dose regions with steep dose gradients and to determine the peak skin dose, without missing hot spots, over all ranges used during interventional radiology procedures. Use of the corset minimized variations associated with angular dependence.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT