Related Experiment Video
Updated: Jul 14, 2026

Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
Evaluation and comparison of two Stuart's liquid swab transport systems tested by the CLSI M40 method
1Department of Pathology, Mt. Sinai Medical Center, One Gustave Levy Place, New York, NY 10029, USA. kvanhorn@focusdx.com
Abstract:
The Copan Diagnostics HealthLink Transporter swab and the Remel Bacti-Swab were evaluated by the Clinical and Laboratory Standards Institute M40-A quantitative swab elution method for maintenance of bacterial viability of six ATCC and six clinical strains. Acceptable levels of viability according to Clinical and Laboratory Standards Institute criteria were maintained for all organisms tested with the HealthLink Transporter swab and for 8 of 12 organisms tested with the Bacti-Swab. In this study, the HealthLink Transporter swab maintained organism viability better than the Bacti-Swab.

