Comparison of performance characteristics of three real-time reverse transcription-PCR test systems for detection and

M Fernanda Sábato1, Mitchell L Shiffman, Michael R Langley

  • 1Molecular Diagnostics Laboratory, Department of Pathology, Virginia Commonwealth University, Richmond, Virginia 23298-0248, USA.

Insights

Two real-time PCR systems, Roche RUO-MPLC and Abbott ASR-Q, effectively detect and quantify hepatitis C virus (HCV) RNA. These systems demonstrated good correlation and agreement with the established COBAS Amplicor HCV Monitor v2.0 assay.

Area of Science:

  • Virology
  • Molecular Diagnostics
  • Infectious Diseases

Background:

  • Hepatitis C virus (HCV) infection remains a significant global health concern.
  • Accurate viral load quantification is crucial for monitoring treatment efficacy and disease progression.
  • Real-time reverse transcription-PCR (RT-PCR) assays are widely used for HCV RNA detection and quantification.

Purpose of the Study:

  • To evaluate the performance characteristics of three real-time RT-PCR test systems for HCV detection and quantification.
  • To directly compare the performance of these systems using the same clinical specimens.
  • To assess the correlation and agreement of the evaluated systems with a reference method.

Main Methods:

  • Performance evaluation of three real-time RT-PCR systems: Roche COBAS TaqMan HCV (RUO-MPLC), Abbott analyte-specific reagents (ASR-Q), and Roche ASR-HP.
  • Assessment of linear range, sensitivity, and precision using commercial HCV panels (genotype 1b).
  • Direct comparison of results with the COBAS Amplicor HCV Monitor v2.0 using 76 clinical specimens with various HCV genotypes and titers.

Main Results:

  • Roche RUO-MPLC and Abbott ASR-Q demonstrated high sensitivity (1.0 log(10) IU/ml) and a broad linear dynamic range (1.0-7.0 log(10) IU/ml).
  • Abbott ASR-Q exhibited the highest reproducibility among the three systems.
  • Good correlation (R(2) = 0.84 for RUO-MPLC, R(2) = 0.93 for ASR-Q) and agreement were observed between Roche RUO-MPLC, Abbott ASR-Q, and the reference method. Roche ASR-HP showed poor correlation and significant underestimation of HCV RNA for certain genotypes.

Conclusions:

  • The Roche RUO-MPLC and Abbott ASR-Q real-time RT-PCR systems provide acceptable performance for HCV RNA detection and quantification.
  • These two systems demonstrate sufficient agreement with the COBAS Amplicor HCV Monitor v2.0, making them suitable for clinical use.
  • The Roche ASR-HP system showed limitations in accuracy and genotype-dependent bias, suggesting it is less reliable for clinical application.