Acetabular fixation options: first-generation modular cup curtain calls and caveats
William G Hamilton1, Cory L Calendine, Sarah E Beykirch
1Anderson Orthopaedic Research Institute, Alexandria, Virginia 22307, USA.
The Journal of Arthroplasty
|August 25, 2007
Summary
This study evaluated three first-generation porous-coated acetabular cups in total hip arthroplasty. The Harris-Galante cups showed significantly better 15-year survivorship compared to Arthopor and ACS Triloc+ designs.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- First-generation porous-coated acetabular cups were widely used in total hip arthroplasty.
- Assessing the long-term outcomes of these early designs is crucial for understanding implant performance.
- Modular hemispheric cups represent a significant advancement in hip implant technology.
Purpose of the Study:
- To review the 15-year clinical outcomes of three distinct types of modular, first-generation, porous-coated acetabular cups.
- To compare the survivorship rates of Arthopor, ACS Triloc+, and Harris-Galante cup designs.
- To identify factors influencing revision rates, including aseptic loosening and polyethylene wear.
Main Methods:
- Retrospective review of 910 primary total hip arthroplasties at a single institution.
- Analysis of implant survivorship using two endpoints: revision for any reason and revision for aseptic loosening.
- Statistical comparison of 15-year survivorship between the three cup designs (Arthropor, ACS Triloc+, Harris-Galante).
Main Results:
- Fifteen-year survivorship (revision for any reason) was 74.9% for Arthopor, 70.1% for ACS Triloc+, and 89.3% for Harris-Galante cups (P = .02).
- Aseptic loosening occurred in 20 of 168 acetabular revisions.
- Survivorship exceeded 95% at 15 years for all designs when using revision for aseptic loosening as the endpoint.
Conclusions:
- The Harris-Galante acetabular cup demonstrated superior 15-year survivorship compared to Arthopor and ACS Triloc+ designs.
- While polyethylene wear and occasional delamination necessitated revisions, aseptic loosening remained low across all three first-generation porous-coated cup types.
- Long-term performance data for early porous-coated acetabular components inform current orthopedic implant development and patient selection.