Digital pulse shape discrimination

L F Miller1, J Preston, S Pozzi

  • 1Department of Nuclear Engineering, The University of Tennessee, Knoxville, Tennessee 37996-2300, USA. lfmiller@utk.edu

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring pulse01:13

Special considerations while measuring pulse

Assessing a patient's pulse is a fundamental skill in healthcare, but certain situations require special attention:
Rectangular and Triangular Pulse Function01:19

Rectangular and Triangular Pulse Function

The unit rectangular pulse function is mathematically represented by a rectangular function centered at the origin with a height of one unit. This function is defined by two parameters: T, which specifies the center location of the pulse along the time axis, and τ, which determines the pulse duration.
For example, consider a rectangular pulse with a 5V amplitude, a 3-second duration, and centered at t=2 seconds. This pulse can be expressed using the rectangular function, written as,
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
Pulse amplitude and quality01:17

Pulse amplitude and quality

Pulse amplitude is a crucial indicator of cardiac health because it provides valuable insights into the strength of left ventricular contractions and the overall uniformity of blood circulation within the vasculature. The strength of the pulse is directly related to the force with which the heart contracts and the volume of blood being pumped.
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...