Validation of a new arterial pulse contour-based cardiac output device
Eric E C de Waal1, Cor J Kalkman, Steffen Rex
1Division of Perioperative and Emergency Care, University Medical Center, Utrecht, The Netherlands. e.e.c.dewaal@azu.nl
Objective:
To evaluate the accuracy and precision of an arterial pulse contour-based continuous cardiac output device (Vigileo). Vigileo cardiac output (VigileoCO) was compared with intermittent transpulmonary thermodilution cardiac output (TPCO) and an established arterial pulse contour-based cardiac output (PCCO).
Design:
Prospective clinical study.
Setting:
University hospital.
Patients:
Twenty-two patients undergoing coronary artery bypass graft surgery.
Interventions:
Defined volume load during surgery and in the postoperative period.
Measurements And Main Results:
We obtained 184 pairs of VigileoCO and TPCO, 140 pairs of VigileoCO and PCCO, and 140 pairs of PCCO and TPCO. Measurements were performed after induction of anesthesia (T1), after sternotomy (T2), immediately after (T3) and 20 mins after volume challenge with 10 mL.kg hydroxyethyl starch 6% (T4), 15 mins after coronary pulmonary bypass (T5), after retransfusion of autologous blood (T6), after arrival at the intensive care unit (T7), and immediately after (T8) and 20 mins after (T9) a second volume load with 10 mL.kg hydroxyethyl starch 6%. TPCO was used to calibrate PCCO. For pooled data, including uncalibrated PCCO data immediately after weaning from coronary pulmonary bypass (T5), the correlation coefficient of TPCO vs. VigileoCO, PCCO vs. VigileoCO, and TPCO vs. PCCO was 0.75, 0.60, and 0.75 respectively. Bland-Altman analysis showed a bias of 0.00, -0.01, and 0.02 L.min, the precision (=sd) was 0.87, 1.08, and 0.93 L.min, and the mean error was 33%, 40%, and 35%. When we compared calibrated PCCO values (T2-T4, T6, T7-9), the correlation coefficients of PCCO-VigileoCO and TPCO-PCCO were 0.72 and 0.85, bias was -0.16 and 0.19 L.min, and mean error was 33% and 27%, respectively. Best correlations and the least differences between TPCO and VigileoCO were observed in postbypass closed-chest conditions and in the intensive care unit.
Conclusions:
Our results showed that VigileoCO enables clinically acceptable assessment of cardiac output in postbypass closed-chest conditions and during stable conditions in the intensive care unit.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: