Related Experiment Video
Updated: Jul 14, 2026

Underwater Endoscopic Injection Sclerotherapy for Gastroesophageal Varices
Published on: August 1, 2025
Accuracy and interobserver agreement of small-caliber vs. conventional esophagogastroduodenoscopy for evaluating
S Pungpapong1, A Keaveny, M Raimondo
1Department of Gastroenterology and Hepatology, Mayo Clinic College of Medicine, Jacksonville, Florida 32224, USA. pungpapong.surakit@mayo.edu
Background And Study Aims:
Advances in endoscope design have allowed high-quality imaging using small-caliber endoscopes (< 6 mm), and these have been proposed as providing an accurate modality for evaluating esophageal varices in several small studies. We aimed to evaluate the accuracy and interobserver agreement of small-caliber esophagogastroduodenoscopy (EGD) compared with conventional EGD for evaluating esophageal varices in a large prospective cohort.
Patients And Methods:
A total of 115 patients with end-stage liver disease and/or portal hypertension were prospectively enrolled into the study. EGD procedures were performed using conventional (8.6-mm) and small-caliber (4.9-mm) endoscopes, back to back and under standard sedation, by two different endoscopists. Esophageal varices were graded at the time of EGD (the "real-time" grade); and by retrospective review of photographs by three endoscopists, when a "consensus" grade (i. e. a grading agreed by two out of the three endoscopists) was used as the final result.
Results:
Of the 115 patients, 33 patients (29 %) were classified as Child's class A, 47 patients (41 %) as Child's class B, and 35 patients (30 %) as Child's class C. The mean model for end-stage liver disease (MELD) score was 13.6. Thirty-six patients (31 %) had undergone previous ligation of esophageal varices. Compared with conventional EGD, the accuracy of small-caliber EGD for esophageal varices grading was 94 % (consensus grade) and 95 % (real-time grade). Excellent concordance was demonstrated between real-time grade and consensus grade, with a kappa of 0.95 for both types of EGD. There was excellent interobserver agreement between endoscopists, regardless of the type of EGD. The severity of hepatic dysfunction and the presence or absence of a history of previous esophageal varices ligation did not have any impact on the accuracy or interobserver agreement.
Conclusions:
Small-caliber EGD performed under sedation via oral intubation is a highly accurate and reliable modality for evaluating esophageal varices in patients with end-stage liver disease and/or portal hypertension, regardless of the degree of hepatic dysfunction or history of previous esophageal varices ligation.
More Related Videos
Related Concept Videos
Esophageal Varices-I: Introduction
Esophageal Varices-II: Clinical Features and Management
In the initial assessment, a thorough review of the patient's medical history is vital to identify risk factors such as liver disease, alcohol abuse, or...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Varicose Veins II: Diagnostic Studies and Interprofessional Care