Resolution of T-wave oversensing with implantable cardioverter defibrillator generator replacement in an adolescent
1Arkansas Children's Hospital, University of Arkansas for Medical Sciences, Little Rock, Arkansas 72202, USA. tuzcuvolkan@uams.edu
Abstract:
Inappropriate delivery of therapy due to T-wave oversensing is a frequent complication of intracardiac cardioverter defibrillators (ICD). A 16-year-old adolescent with congenital long QT syndrome type 3 underwent single-chamber implantable cardioverter defibrillator (ICD) (EnTrust VR, Medtronic Inc., Minneapolis, MN, USA) implantation.He received two appropriate and four inappropriate shocks. The inappropriate shocks were due to T-wave oversensing. Following the replacement of ICD with a different brand generator (Vitality 2 VR, Boston Scientific Inc., Natick, MA, USA), no more T-wave oversensing is noted. Proper filtering is imperative in ICDs in order to prevent T-wave oversensing-related inappropriate delivery of therapy.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Erikson's Theory on Socioemotional Development during Adolescence
Continuous Renal Replacement Therapy
Cognitive Development During Adolescence
The Wave Nature of Light
Revisionist Views of Adolescent and Adult Cognition
Wave Parameters