Experimental study on shear stress distributions in a centrifugal blood pump
Hiroshi Mizunuma1, Ryou Nakajima
1Department of Mechanical Engineering, Tokyo Metropolitan University, Tokyo, Japan. mizunuma@ecomp.metro-u.ac.jp
Abstract:
Wall shear stress on the pump casing cover was measured using a surface-mounted hot-film sensor. In addition, the shear stress distribution in the pump was qualitatively investigated by means of oil-film visualization. The characteristics of shear stress in the pump are discussed, including the results on the oil-film visualization. The centrifugal blood pump used was a Nikkiso HPM-15 (Nikkiso Co., Ltd, Tokyo, Japan). The hot-film measurement indicated that the shear stress was approximately proportional to the rotating speed, and exceeded 300 Pa when r/R > 0.5 at 3000 rpm. The circumferential average shear stress on the casing cover was of the same order as the characteristic stress sigma obtained from the pump axial torque. These results suggest that the shear stress on the casing cover can be used to evaluate the characteristic shear stress in the pump.
More Related Videos
Related Concept Videos
Stress Concentrations in Circular Shafts
Thin-Walled Hollow Shafts
Steady, Laminar Flow Between Parallel Plates
Design of Transmission Shafts - Stress Analysis
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Newtonian Fluid: Problem Solving
A velocity gradient forms within the fluid when a Newtonian fluid is placed between two parallel plates, with...