Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Atmospheric pressure ionization permanent magnet fourier transform ion cyclotron resonance mass spectrometry
Andrey N Vilkov1, Chaminda M Gamage, Alexander S Misharin
1MassTech, Inc., Columbia, Maryland 21046, USA. andrey_vilkov@apmaldi.com
Abstract:
A new Fourier transform ion cyclotron resonance mass spectrometer based on a permanent magnet with an atmospheric pressure ionization source was designed and constructed. A mass resolving power (full-width-at-half-maximum) of up to 80,000 in the electron ionization mode and 25,000 in the electrospray mode was obtained. Also, a mass measurement accuracy at low-ppm level has been demonstrated for peptide mixtures in a mass range of up to 1200 m/z in the isotopically resolved mass spectra.