Impact of creatinine calibration on performance of GFR estimating equations in a pooled individual patient database
Lesley A Stevens1, Jane Manzi, Andrew S Levey
1Division of Nephrology, Tufts-New England Medical Center, 750 Washington St, Box #391, Boston, MA 02111, USA. lstevens1@tufts-nemc.org <lstevens1@tufts-nemc.org>
Background:
Variation in performance of glomerular filtration rate (GFR) estimating equations is related to variation in calibration of the creatinine assay across clinical laboratories.
Study Design:
Cross-sectional analysis.
Setting & Participants:
6 research studies and 4 clinical populations including 5,504 participants who had GFR measured using urinary clearance of iothalamate.
Measurements:
Standardized serum creatinine values obtained by means of calibration to the Cleveland Clinic Research Laboratory using frozen specimens, a calibration panel, and/or survey results from the College of American Pathologists.
Predictor:
Noncalibrated serum creatinine assayed in research and clinical laboratories compared with standardized serum creatinine.
Outcome:
Difference between measured GFR versus GFR estimated from the Modification of Diet in Renal Disease (MDRD) Study and Cockcroft-Gault equations.
Results:
For a noncalibrated serum creatinine value of 1 mg/dL (88.4 micromol/L), standardized serum creatinine value was 0.07 mg/dL (6.2 micromol/L) less than noncalibrated values. In the pooled data set, for the MDRD Study equation, calibration improved median percentage of difference between measured and estimated GFR from 9.0% (interquartile range [IQR], 28%) to 5.8% (IQR, 28%) and improved the percentage of estimates within 30% of measured GFR (P30) from 80% to 83%. The effect of calibration was greater at higher levels of GFR and varied across studies. For the Cockcroft-Gault equation, calibration worsened the median percentage of difference from -2.0% (IQR, 38%) to -11.4% (IQR, 39%), and the P30, from 74% to 69%.
Limitations:
College of American Pathologist samples were used for calibration of clinical populations; calibration factors do not account for drift over time in the serum creatinine assay; calibration cannot account for variation in assay performance among individuals.
Conclusion:
Calibration improves the performance of the MDRD Study equation. After calibration, larger errors remain for GFR estimates greater than 60 mL/min/1.73 m2 (>1 mL/s/1.73 m2).
Related Concept Videos
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Pharmacokinetics in Geriatric Patients: Effect of Age on Drug Excretion
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...