CT angiography of renal artery anatomy for evaluating embolic protection devices
Mallik R Thatipelli1, Enrique A Sabater, Haraldur Bjarnason
1Department of Radiology, Mayo Clinic, Rochester, MN 55905, USA.
Purpose:
To determine which commercially available embolic protection devices are suitable for use in renal arteries.
Material And Methods:
A retrospective study was conducted of 97 patients who underwent three-dimensional computed tomographic (CT) angiography with either 16- or 64-detector row helical CT of the renal arteries for the evaluation of renal artery stenosis. Nine anatomic measurements were obtained of the renal artery from each reformatted CT angiogram. These data were used to evaluate seven commercially available embolic protection devices (SpideRX, BSC FilterWire EZ, GuardWire, Angioguard, Accunet, Emboshield, TriActiv FX) to determine suitability for use in conjunction with renal artery stent placement. The authors evaluated the reference vessel diameter (diameter to which the stent would be expanded) and length of device.
Results:
The mean length (+/- standard deviation) of the main renal artery on the left (39.9 mm +/- 6.7) was significantly less than that on the right (44.9 mm +/- 7.4, P < .001). The mean distal diameter of the main renal artery was significantly larger on the left (6.0 mm +/- 0.5) than on the right (5.6 mm +/- 0.6, P = .02). For a reference vessel diameter of 4-7 mm necessitating a 12-mm-long stent, only three of the seven devices tested (SpideRx, Angioguard, and Accunet) could be used.
Conclusion:
Three of the seven devices tested could be used for reference vessel diameters of 4-7 mm necessitating a 12-mm-long stent.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...