Measurement of hold-up volumes in reverse-phase liquid chromatography Definition and comparison between static and
Fabrice Gritti1, Yuri Kazakevich, Georges Guiochon
1Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:
The hold-up volumes, V(M) of two series of RPLC adsorbents were measured using three different approaches. The first method is based on the difference between the volumes of the empty column tube (150x4.6mm) and of the material packed inside the column. It is considered as giving the correct value of V(M). This method combines the results of the BET characterization of the adsorbent before packing (giving the specific pore volume), of carbon element analysis (giving the mass fraction of silica and alkyl bonded chains), of Helium pycnometry (providing silica density), and of inverse size exclusion chromatography (ISEC) performed on the packed column (yielding the interparticle volume). The second method is static pycnometry, which consists in weighing the masses of the chromatographic column filled with two distinct solvents of different densities. The last method is based on the thermodynamic definition of the hold-up volume and uses the dynamic minor disturbance method (MDM) with binary eluents. The experimental results of these three non-destructive methods are compared. They exhibit significant, systematic differences. Pycnometry underestimates V(M) by a few percent for adsorbents having a high carbon content. The results of the MDM method depend strongly on the choice of the binary solution used and may underestimate or overestimate V(M). The hold-up volume V(M) of the RPLC adsorbents tested is best measured by the MDM method using a mixture of ethanol and water.
More Related Videos
11:38Combining Microfluidics and Microrheology to Determine Rheological Properties of Soft Matter during Repeated Phase Transitions
Published on: April 19, 2018
11:25Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Related Concept Videos
Chromatographic Methods: Terminology
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...