Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient
Ryoko Tsuchida1, Kazuyuki Araki, Tomohiro Okano
1Department of Radiology, Showa University School of Medicine, Tokyo, Japan.
Objectives:
To assess the accuracy of a limited cone-beam volumetric imaging system in detecting incipient proximal caries.
Study Design:
Fifty extracted premolars were used. On 100 surfaces examined, 71 proximal carious lesions were found; 47 were limited to the enamel, and 24 penetrated into the dentin. A limited cone-beam volumetric imaging system (CBVI), 3D Accuitomo (J. Morita Mfg. Corp., Kyoto, Japan), was used. Insight films (Eastman Kodak, Rochester, NY) were used for comparison. Seven experienced oral radiologists determined the presence of lesions using a continuous rating scale. The area measure (Az) values were calculated from the receiver operating characteristic curve obtained for each modality and observer.
Results:
The mean Az-values were 0.63 +/- 0.02 for 3D Accuitomo and 0.63 +/- 0.03 for film. Analysis of variance showed no statistically significant difference between 3D Accuitomo and film.
Conclusions:
A limited CBVI, 3D Accuitomo, could not enhance the accuracy in detecting proximal carious lesions.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...