Related Experiment Video
Updated: Jul 13, 2026

An Improved Technique for Trimethylamine Detection in Animal-Derived Medicine by Headspace Gas Chromatography-Tandem Quadrupole Mass Spectrometry
Published on: March 10, 2023
Determination of formaldehyde in urine by headspace gas chromatography
Akito Takeuchi1, Tomoko Takigawa, Mikiko Abe
1Department of Public Health, Okayama University, Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, 2-5-1, Shikata-cho, Okayama, 700-8558, Japan.
Abstract:
Formaldehyde is a carcinogen to which humans are exposed daily, but few methods are available to quantify formaldehyde in biological samples. We developed a simple, sensitive and rapid technique for the quantification of formaldehyde in urine by derivatization with O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine, using a headspace sampler coupled to a gas chromatograph equipped with an electron capture detector. The detection limit was 1.08 microg/L. The overall recovery of formaldehyde spiked in urine was 99%. The concentration of formaldehyde in urine obtained from healthy volunteers ranged from 56.85 to 144.57 microg/L. This method can be used successfully to measure formaldehyde in urine.