Related Experiment Video
Updated: Jul 13, 2026

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing
Sebastian Leschka1, Hans Scheffel, Lotus Desbiolles
1Institute of Diagnostic Radiology, University Hospital Zurich, Zurich, Switzerland.
Purpose:
To evaluate reconstruction intervals and image quality in dual-source computed tomography (DSCT) coronary angiography for optimal placement of the ECG-pulsing window.
Materials And Methods:
DSCT coronary angiography was performed in 60 patients. Thirteen datasets were reconstructed in 5% increments from 20-80%. Two readers independently assessed image quality of each segment in each percentage-interval, using scores ranging from 1 (no motion artifacts) to 4 (nonevaluable).
Results:
Mean heart rate (HR) was 69.0 +/- 18.9 beats per minute (bpm) (range, 35-117 bpm). Diagnostic image quality (scores 1-3) was found in 97.8% of all segments (763 of 780). The 70% RR-interval provided best image quality in all patients and all HRs. The narrowest reconstruction window providing diagnostic image quality was 60-70% for HR <60 bpm, 60-80% for 60-70 bpm, 55-80% for 70-80 bpm, and 30-80% for HR >80 bpm.
Conclusions:
DSCT coronary angiography provides best image quality for various HRs at 70%. The ECG-pulsing window can be adapted according to the HR while maintaining diagnostic image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...