Quantitative determination of BMS-378806 in human plasma and urine by high-performance liquid chromatography/tandem
Y J Xue1, Jing-He Yan, Mark Arnold
1Pharmaceutical Candidate Optimization, Pharmaceutical Research Institute, Bristol-Myers Squibb, New Brunswick, NJ 08903, USA. yongjun.xue@bms.com
Abstract:
BMS-378806 is a human immunodeficiency virus (HIV) entry inhibitor that is being developed for the oral treatment of HIV infection. Human plasma and urine LC/MS/ MS methods have been developed and validated for the quantitation of BMS-378806. For human plasma method, methyl t-butyl ether was used to extract BMS-378806 from plasma in a 96-well format, and the organic layers were dried down and then reconstituted for the injection, while a dilute-and-shoot approach was used for human urine method in a 96-well format. Chromatographic separation was achieved isocratically on a Phenomenex C18 (2) Luna column (2 x 50 mm2, 5 microm). The mobile phase contained 60:40 v/v of 0.1% formic acid in water and ACN. Detection was by positive ion electrospray MS/MS. The standard curves ranged from 1.25 to 1000 ng/mL for the plasma assay and from 10 to 5000 ng/mL for the urine assay. The curves were fitted to a 1/x2 weighted quadratic regression model for both methods. The validation results demonstrated that both methods had satisfactory precision and accuracy across the calibration ranges. The methods were applied to the analysis of human plasma and urine samples from a single ascending dose clinical study to assess the pharmacokinetics of the drug. The pharmacokinetic analysis results indicated the absorption and disposition of the drug was rapid. The systemic exposure of BMS-378806 was generally dose proportional among the doses from 100 to 1200 mg, but not dose proportional to 1600 mg. There were modest increases in the systemic exposure when the drug was given with food or given as a solution formulation. Renal excretion was not a substantial elimination pathway of the drug. BMS378806 was safe and well tolerated over a dose range of 100-1600 mg administered as a single oral dose.
More Related Videos
11:25Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Types of Detectors